Transducer Position & LV Geometry

[Page 4]

Erroneous calculation of left ventricular volumes and ejection fraction when the long axis is foreshortened

Figure 9. An example of the consequences of LV foreshortening.
In this patient, foreshortening is prevalent in systole, the LV long axis is foreshortened from 10.6 cm (right panel) to 7.6 cm (left panel), which produces a great overestimation of EF= 66.2 % (overestimated) vs 43.2 % (correct value).
The clinical consequence of this incorrect evaluation was that the patient was withheld from ACE-inhibitors for 2 years.

Incorrect position of the transducer on the thorax anterior to the anterior axillary line
Figure 10
Correct position of the transducer on the thorax along the mid-axillary line
Figure 11

The correct transducer position to obtain the true LV long axis in the 4-chamber view is very often much more lateral than initially imagined by the operator. Figure 11 shows the correct position of the transducer along the mid-axillary line, whereas in Figure 10 the transducer is erroneously placed anterior to the anterior axillary line. The position of the transducer depends obviously on the alignment of the heart within the thorax: when the heart is positioned vertically (Figure 12), then the correct transducer position is similar to that shown in Figure 10. In contrast, when the heart is more horizontal (Figure 13), the correct transducer position is at the mid or posterior axillary line (Figure 11).

Simulated transducer positioning on chest Xray in a subject with the LV long axis parallel to the spine
Figure 12. Heart with the LV long axis more parallel to the spine
Simulated transducer positioning on chest Xray in a subject with LV long axis more parallel to the diaphragm
Figure 13. Heart with LV long axis more parallel to the diaphragm

Leave a Reply